Home > News & Information >

News & Information

The daily use and maintenance of salt spray test chamber, maintenance of salt spray test c



1.Please confirm the power connection before equipment operation
2.Please confirm the automatic water supply system
3.Please confirm the air supply system
4.Please confirm the tank is sealed
5.Vent inspection
6.Please use analytical pure NaCl and distilled or deionized water when preparing salt solutions, and use them immediately.
7.After each test, the power supply, gas source and water source should be cut off to avoid the equipment being in standby mode for a long time.
8.At the end of each test, it is recommended to clean the test chamber (including spray chamber, salt solution chamber, preheating tank and sealing tank) with clean water to keep the equipment clean
9.In each test or at the end of the test, the solution of the standard measuring cup should be emptied and cleaned in time to avoid the accumulation of crystals in the salt solution, which will affect the settlement calculation
10.When cleaning the inside of the chamber,please note:
The protection of temperature sensor's protective layer
The protection of glass filters and glass nozzles (do not use needles or any hard objects to unclog the filters or nozzles)

Model number* Dimension of Working room* Overall Dimension* Power* Voltage*
LRHS-108-RY 450×600×400 780×1140×1050 1.6kW 220V
LRHS-270-RY 600×900×500 960×1460×1150 2.5kW 220V
LRHS-412-RY 750×1100×500 1100×1730×1240 3.0kW 220V
LRHS-663-RY 850×1300×600 1230×2150×1410 5.6kW 380V
LRHS-816-RY 850×1600×600 1230×2450×1460 6.5kW 380V
LRHS-1080-RY 900×2000×600 1280×2850×1460 6.7kW 380V
 
Technical parameters
Laboratory temperature:RT+5℃-50℃(Set according to the test standard)
Material:Glass fiber reinforced plastic (FRP)
Heater:Nickel chromium wire heater
Mist trap:YES
Temperature sensor:PT100 platinum resistance
Temperature controller:Imported digital display instrument, digital setting /PID automatic adjustment
Temperature fluctuation:±0.5℃
Salt fog fall-out rate:1-2ml/h.80cm2
Spray tower:Acrylic conical disperser designed by ourselves can adjust the spray size and balance the fog
Spray method:Continuous, gap spray optional
 
 
Attachment:Fog bank, water tank, fog collector, support rack
Safety device:Over temperature protection, leakage protection
Note:No special requirements can be customized

More details, please contact sales@lenpure.com