How to solve salt spray chamber stop to spray

Salt Spray Test Machine is used to evaluate the anti-corrosion ability of all the materials surfaces of coating,layer, electroplating in condition of salt spray environment.

Our standard models meet the requirements of basic, continuous salt spray tests conducted at a single temperature only, such as ASTM B117 and similar international test standards.

It is comply with Netural salt fog test(NSS)

Salt spray test(SS)

Acetic Acid Salt Spray Test(ASS)

Copper Accelerated Acetic Acid Salt Spray Test(CASS)

 

In the course of the salt spray fog test, sometimes there will be some problems that cannot reach the spray state. In order to achieve an ideal salt spray test also need to understanding of the following:

When the fog in the salt spray chamber is insufficient -- Reason & Solution:

1) Spray regulator placed too lower, should adjust the height of the spray regulator;

2) The glass filter in the preheating tank was clogged, and the filter should be cleaned regularly in this case;

3) Equipment pressure setting is too low, this phenomenon needs to adjust the control valve, adjust to 1Kg / ㎝ 2 pressure level. At the same time, the pressure regulator on the air Compressor needs to adjust to 2Kg / ㎝ 2 pressure level.

 

 

When the salt spray chamber cannot spray, or the nozzle has been blocked -- Reason & Solution:

1) Air compressor stop running, turn on air compressor;

2) The air compressor outlet switch turn-off, turn on it

3) Malfunction of the solenoid valve, pressure gauge failure or the pressure is too low, contact with LINPIN supplier to repair;

4) When the nozzle has been clogged, it is necessary to carefully remove the nozzle to cleaning.