A split structure can be adopted, with the refrigeration unit and the box placed separately;
The top of the box is perforated for easy lifting of the motor;
Box bottom has 4 support holes sealed with molded silicone plugs to prevent frost & water overflow.
The central axis adopts a special sealing device to avoid frosting at low temperatures;
Model | Studio Size (D*W*H)mm | Dimension(D*W*H)mm | |
LRHS-101-NDO₃ | 450×450×500 | 1160×1000×1800 | |
LRHS-225-NDO₃ | 500×600×750 | 1220×1150×2040 | |
LRHS-504-NDO₃ | 700×800×900 | 1520×1250×2200 | |
LRHS-800-NDO₃ | 800×1000×1000 | 1520×1540×2300 | |
LRHS-1000-NDO₃ | 1000×1000×1000 | 1720×1540×2300 |
temperature range | 25℃~65℃(adjustable) | ||||||
Humidity range | 40~98%RH | ||||||
Temperature uniformity | ≤2℃(at no load) | ||||||
Temperature fluctuation | ±0.5℃(at no load) | ||||||
temperature deviation | ±2℃ | ||||||
Humidity Deviation | ±5% RH at ≤75% RH, +2, -3% RH at >75% RH | ||||||
ozone concentration | 50~1000pphm or 50~300ppm | ||||||
gas velocity | 8~16mm/s | ||||||
Ozone concentration deviation | ≤±10% | ||||||
Rotation speed of the sample frame | 1~3r/min(adjustable) | ||||||
Outer case material | High quality A3 steel plate electrostatic spraying | ||||||
Inner box material | Imported high-grade stainless steel SUS304 | ||||||
Insulation material | Rigid polyurethane foam + fiberglass | ||||||
viewing window | 210 x 275mm or 395 x 395mm (effective field of view) | ||||||
Temperature and humidity controller | 7.0-inch color touch screen of “Taiwan Weilun” and PLC module control of “Germany Siemens”. | ||||||
Temperature and humidity sensors | PT100 platinum resistance temperature measurement body | ||||||
Ozone Sensor | ECO USA | ||||||
Cooling method | stand-alone refrigeration | ||||||
Compressor | Taikang, France | ||||||
Refrigerant | R404A (environmentally friendly) (ozone depletion index 0) | ||||||
standard configuration | Carousel type sample holder 1 set | ||||||
security protection | Compressor overpressure overload, heating tube dry burning, water supply abnormality, fan motor overheating, equipment over temperature Overall equipment under phase/reverse phase, test timing, leakage protection, automatic shutdown after the fault alarm and other protections |
||||||
Supply Voltage | AC380V±10% 50Hz | ||||||
Power | 4.0kW/5.5kW/7.0kW/9.0kW/11.5kW | ||||||
operating environment | 5℃~+28℃ ≤85% RH | ||||||
Note. 1: 1、“LRHS-101-NDO₃” for dynamic ozone aging test chamber, static ozone aging test model “LRHS-101-NO₃”. 2、Can be customized according to the user’s specific test requirements of non-standard ozone aging test chamber. 3、This technical information, subject to change without notice |